MSS Chemilume CL100

MSS Chemilume CL400
[Click picture to enlarge]

 

DIN 51835-1   Oxidation stability of oils   tl_files/mss/images/blue.gif DIN


DIN 51835-2   Oxidation stability of lubricating greases   tl_files/mss/images/blue.gif DIN

 


The fundamental cause for the decomposition of most organic material is oxidation reaction. Oxidation in the presence of oxygen happens under common environmental condition whilst treatment and application. For example under higher temperatures and influence of UV-radiation. Oxidation mostly leads to a negative alteration of the material properties.
Therefore the oxidation stability is one of the essential problems during treatment as well as during long term usage of polymeric materials. As consequence the labeling of the oxidation stability of material is one of the essential targets of test technique.
The decomposition of polymeric material can be caused by different factors like mechanical deformation, heat, UV-light, ionizing radiation and others. In general the fact is accepted, that all these factors initiate a reaction with oxygen within the polymers, which is the main cause for the decomposition. The circumstance that the oxygen reaction of polymers was accompanied by small light emissions (Chemiluminescence) encouraged the development of a new CL-sensitive technique, which turned out to be easier and supply information, that were not available by the conventional measuring methods (Oven tests, Infrared-Spectrometry (IR), Oxygen Absorption and Differential-Scanning-Calometry (DSC)).
 
Innovative application of modern technique accelerates the material testing.

Chemiluminescence and Oxidation

Most organic material is decomposited by autocatalytic oxidation that is initiated by influence of heat during treatment (e.g. during extrusion) resp. during intended use or by influence of UV-radiation. During oxidation hydro peroxides are generated which emit light during decomposition. This light is detected and a measure for the grade of oxidation.

The Chemiluminescence Signal

To test the oxidation stability the probe is heated in an inert atmosphere and then exposed to pure oxygen. The oxidation increases to an autocatalytic stationary state. The induction period and the velocity constant for this stationary state are a measure for the materials oxidation resistance. With the CL it is often possible to differentiate signals between probes where other methods like determination of Oxygen-Induction Time (OIT) by Differential Scanning Calorimetry (DSC) or the oxygen absorption fail.
 

ChemiLume-Software

By means of the powerful, menu-driven software setting of test parameters, instrument control, data acquisition, graphical charts and quantitative data analysis is easy.

Versatile Test Possibilities

By isotherm tests in oxygen the thermal oxidation stability can be analyzed. By means of linear temperature rise in nitrogen an already existing material oxidation can be analyzed.

Thermal Stability of Petroleum Products

Chemiluminesczence requires a material amount of only some milligrams as pellets, flakes, powder or liquid probes. During formulation of natural and synthetic engine oil, which in some extends contained anti-oxidants, differences in induction periods and oxidation rate constants were found. The rating of these materials was done by a lifespan, defined as the ratio of induction period (dimensionless) and oxidation rate constant.
Mineral oil had the lowest, purged oil with vitamin E the highest oxidation stability.
 

Chemiluminescence for Stability Tests

The Chemiluminescence can be used for the evaluation of thermal and photo oxidative stability of untreated and stabilized polymers, the effectiveness of additives and the impact of treatment- or operating-temperatures, storage conditions as well as influences of weather.
It is also used for organic material like food, drugs and biological material.
 

Testing with Light Speed

Polypropylene for commercial dishwashers was stabilized by different additives and concentrations. By means of Chemiluminescence an evaluation of stability was possible in the same rank and dimension as with oven aging tests, but within a fraction of the time. In contrast to subjective evaluations with the oven aging test a quantitative analysis of the material stability was possible.
Chemiluminescence tests allow a fast evaluation which materials are best applicable for certain applications. Stability tests on polymers are done in oxygen at temperatures that are 20 to 50° C below the treatment temperature or during melting-index-test used temperatures. In the test cells precise temperatures between room temperature and 250° C can be set. A cell conditioning at 300° C minimizes the use of cleaning agens.
With the MSS Chemilume CL 100 the scientist for polymers, additives and organic material has a powerful analytic tool available. It is robust, reliable and user friendly within the daily control of product quality are material selection.
 

Testing with the CL 100

The MSS Chemilume CL 100 contains one test cell. The delivered software supplies detailed data concerning the CL-signal and enables a quantitative and qualitative analysis of the CL-curve.

MSS Chemilume Standard Equipment

- Probe carrier made of 99,5% pure aluminum
- Photon counter-Photomultiplier
- Gas flow control
- Maximum test temperature 250° C
- Standard PC with Intel/AMD-Processor
- Windows2000 and WindowsXP-supported software
- Data acquisition
- Probe forceps

MSS Chemilume Specifications

 
Need for cooled water
0,2 l/min.
Water pressure
70 kPa
Temperature range
16 - 21° C
Maximum test temperature
250° C
Need for gas
Oxygen and Nitrogen
Gas flow
25 or 50 ml/Min. /cell
Optical sensitivity
Approx. 2 Counts /sec.
   
Dimensions:
 
Widths / Heights / Depths
52cm / 31cm / 51cm
Weight
38 kg
   
Electrical Spezifications:
115- 230V A/C, 360 Watt (max.)



















MSS Chemilume Accessories


Circular Cooler

Supplies water cooled to constant temperature to cool the photomultiplier.
100/200/230 V A/C, 50 Hz or 120 V A/C, 60 Hz.

Precision Pressure Controller

Second stage, 70 ­ 175 kPa for O2 as well as for N2.

Photomultiplier-Calibration Kit

Contains light standard, neutral grey filter and calibration holder.
Will be provided free of charge on loan.

Temperature-Calibration Kit

Contains a NIST-standard traced temperature calibration device and thermo element.

Oil Test Kit

Contains pot made of  99,5% pure aluminum and cover glasses for volatile, liquid probes.


MSS Chemilume CL400 Prospect MSS Chemilume CL400 Prospect


PP-film at 180°C
MSS Chemilume CL400
[Click picture to enlarge]



::Top::
Subject to technical changes.